Abstract
A prototype blackbody radiation source has been developed which incorporates a double cone cavity of extremely uniform surface temperature. This uniformity of surface temperature is the result of a unique method of supplying heat to the cavity. The principle of operation is based on heat pipe techniques which transfer heat by change of phase and mass transfer. The cavity is heated by serving as the condenser section of the heat pipe which is inherently a nearly isothermal device. Optical tests show that the isothermal cavity has a lambertian distribution over an inclusive angle of better than 60 degrees while operating between 419 degrees C and 760 degrees C.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.