Abstract
This document, from the International Society for Clinical Electrophysiology of Vision (ISCEV), presents an updated and revised ISCEV Standard for full-field clinical electroretinography (ffERG or simply ERG). The parameters for Standard flash stimuli have been revised to accommodate a variety of light sources including gas discharge lamps and light emitting diodes. This ISCEV Standard for clinical ERGs specifies six responses based on the adaptation state of the eye and the flash strength: (1) Dark-adapted 0.01 ERG (rod ERG); (2) Dark-adapted 3 ERG (combined rod-cone standard flash ERG); (3) Dark-adapted 3 oscillatory potentials; (4) Dark-adapted 10 ERG (strong flash ERG); (5) Light-adapted 3 ERG (standard flash "cone" ERG); and (6) Light-adapted 30 Hz flicker ERG. ISCEV encourages the use of additional ERG protocols for testing beyond this minimum standard for clinical ERGs.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.