Abstract

In a combined experimental and Monte Carlo study the possible influence of the surrounding material on the response of the alanine dosimetry system was investigated. The aim of this work was to estimate the uncertainties induced by the surroundings with respect to quality assurance measurements for radiotherapy, for example in humanoid phantoms. Six different materials were tested. The electron density range covered comprises the range present in human tissue. No significant influence of the surrounding material could be found for irradiations in the 60Co reference field of the Physikalisch–Technische Bundesanstalt (PTB).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.