Abstract
Background/Aims:Access blood water flow rate (Qaw) can be measured during hemodialysis using an online effective ionic dialysance (EID) methodology. Fresenius employ this methodology in their 2008K dialysis machine. The machine computer converts Qaw to an access blood flow rate (Fresenius Qa) using a generic blood water constant (BWC). We wished to validate this BWC. Methods: 18 patients had Fresenius Qa measurements using the EID and these were compared with a ‘gold standard’ ultrasound dilution methodology (Transonic Qa). Qa values were also obtained by removing the BWC from Fresenius Qa values to obtain the Qaw and recorrecting it with individualized patient factors using hematocrit and total protein values (HctTp Qa). The measurements were repeated after 1 h. Results:There were no significant differences between Fresenius and Transonic, nor between HctTp and Transonic Qa values (p > 0.17). There were strong correlations between both sets of values (r > 0.856; p < 0.001). There was a significant correlation between the pairs of Transonic Qa values (r = 0.823; p < 0.007), but not for Fresenius Qa pairs (r = 0.573; p > 0.07). It was surmised that the BWC was not valid post-dialysis. Conclusion:The generic BWC is comparable to individualized blood water correction factors when Qa measures are made early in dialysis and prior to ultrafiltration treatment.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.