Abstract

Nowadays, many efforts are devoted to improve analytical methods regarding efficiency, analysis time and greenness. In this context, Supercritical Fluid Chromatography (SFC) is often regarded as a good alternative over Normal Phase Liquid Chromatography (NPLC). Indeed, modern SFC separations are fast, efficient with suitable quantitative performances. Moreover, the hyphenation of SFC to mass spectrometry (MS) provides additional gains in specificity and sensitivity. The present work aims at the determination of vitamin D3 by SFC-MS for routine Quality Control (QC) of medicines specifically. Based on the chromatographic parameters previously defined in SFC-UV by Design of Experiments (DoE) and Design Space methodology, the method was adapted to work under isopycnic conditions ensuring a baseline separation of the compounds. Afterwards, the response provided by the MS detector was optimized by means of DoE methodology associated to desirability functions. Using these optimal MS parameters, quantitative performances of the SFC-MS method were challenged by means of total error approach method validation. The resulting accuracy profile demonstrated the full validity of the SFC-MS method. It was indeed possible to meet the specification established by the European Medicines Agency (EMA) (i.e. 95.0 − 105.0% of the API content) for a dosing range corresponding to at least 70.0-130.0% of the API content. These results highlight the possibility to use SFC-MS for the QC of medicine and obviously support the switch to greener analytical methods.