Abstract

Sacral nerve modulation (SNM) is recommended as a first-line surgical therapy for patients with faecal incontinence (FI). During patient follow-up, it is recommended that stimulation settings be reprogrammed to optimize patient outcomes. The aim of the present study was to evaluate the efficacy of stimulator reprogramming in patients with an implanted SNM device to treat FI. The data from patients who received a permanent SNM implant in a single centre from January 2008 to December 2019 were retrospectively analysed. Symptoms that occurred after implantation, the stimulator settings of the SNM device and changes made at each follow-up visit were noted. The efficacy of reprogramming was determined by assessing patient satisfaction. Of the 117 patients (male/female 4/113; mean age 59.5±11.8years) with a SNM implant for FI, 84 (72%) had at least one symptom requiring reprogramming of the stimulator, most often during the first year after implantation (p=0.05). The most frequently reported symptoms were loss of efficacy (68.5%; p=1×10-3 ) and pain (20.5%; p=1×10-3 ). Reprogramming was effective 53% of the time when treating loss of efficacy and 76% of the time when treating pain. When the stimulation parameters were reprogrammed at least four consecutive times to correct a symptom, the reprogramming was less effective in treating the symptom (p=0.02). Regular follow-up of patients with SNM device implants associated with reprogramming of stimulation parameters to improve the treatment of reported symptoms would optimize the efficacy of SNM.