Abstract

Pulmonary valve replacement (PVR) with right ventricular outflow tract (RVOT) reconstruction is a common congenital cardiac operation. Porcine submucosal intestinal-derived extracellular matrix (ECM) patches have been used for RVOT reconstruction. We present 2 adult patients with Tetralogy of Fallot who underwent PVR with RVOT reconstruction utilizing ECM. Both cases required reoperation due to patch dehiscence causing a large paravalvular leak. One patient also had a pseudoaneurysm associated with ECM dehiscence. There may be a propensity for ECM dehiscence in this application and, based on these cases, we recommend avoidance of ECM in RVOT reconstruction with PVR. PVR patients repaired with ECM should be monitored for this complication.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.