The aim of this study was to test the hypothesis that grey levels are a suitable alternative measure of sensitivity in ultrasound imaging quality assurance, as there are several caveats in the use of penetration depth. In a primary cohort of nine probes, where measurements had been made for 6 to 34 mo, both penetration depth and mean grey level fell below tolerance for six probes; both penetration depth and mean grey level remained within tolerance for three probes. In a secondary cohort where a measurement programme had been in place for a shorter period, grey level and/or penetration depth fell below tolerance in 15 of 66 probes; the sensitivity and specificity of at least 10% loss of grey level in predicting >5% loss in penetration depth were 91% and 93%, respectively. A loss of grey level accompanies a loss of penetration and provides a suitable alternative measure of sensitivity.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.