This study followed blood pressure in 11 subjects 5 years after slight or established hypertension had been revealed using a cuff of the correct width (appropriate to arm circumference); at that time, the use of a standard-width cuff (12 cm) had shown a blood pressure within the normal range. For the present study, blood pressure was determined indirectly under very strict conditions so that a faithful comparison between the readings with both cuffs could be achieved. An important underestimate of blood pressure was detected in thin arms when the standard width cuff was used. The present measurements showed a consistent increase in blood pressure with standard-width cuff readings, confirming the hypertension that had been identified using the correct cuff width 5 years previously. These results support our previous hypothesis that early diagnosis of hypertension can be masked by the standard-width cuff, particularly in lean persons.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.