Abstract

Heart rate changes during determination of coronary flow by parametric imaging may influence the flow measurement. Thus, the question is whether atrial pacing is mandatory for determination of coronary flow reserve (CFR) by this technique. CFR was calculated by digital subtraction angiography (parametric imaging) in 10 patients (8 with coronary artery disease and 2 control subjects) during sinus rhythm and during atrial pacing. Flow measurements were determined in the perfusion region of the left anterior descending and circumflex coronary artery, both at rest and after maximal coronary vasodilation with 10 mg intracoronary papaverine. CFR was defined as coronary flow during hyperemia divided by coronary flow at rest. Spontaneous heart rate was 71 ± 15 min −1 at baseline, 73 ± 15 min −1 after papaverine injection and 85 ± 10 min −1 during atrial pacing. Heart rate variations during coronary arteriography were 4 ± 3 min −1 at baseline and 5 ± 4 min −1 after papaverine administration. CFR was 2.61 ± 1.01 during sinus rhythm and 2.67 ± 1.05 during atrial pacing. Mean absolute difference in CFR between sinus rhythm and atrial pacing was 0.31 ± 0.31 (12 ± 10% of CFR during pacing). Spontaneous heart rate variations during coronary arteriography are not associated with significant changes in CFR. Thus, atrial pacing is not mandatory for the determination of CFR by parametric imaging.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call