Abstract
Photodiodes are used as easy-to-operate detectors in the extreme-ultraviolet spectral range. At the Physikalisch-Technische Bundesanstalt photodiodes are calibrated with an uncertainty of spectral responsivity of 0.3% or less. Stable photodiodes are a prerequisite for the dissemination of these high-accuracy calibrations to customers. Silicon photodiodes with different top layers were exposed to intense extreme-ultraviolet irradiation. Diodes coated with diamondlike carbon or TiSiN proved to be stable within a few percent up to a radiant exposure of 100 kJ/cm2. The changes in responsivity could be explained as being due to carbon contamination and to changes in the internal charge collection efficiency. In ultrahigh vacuum, no indication of oxidation was found.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.