Abstract
Within the EURopean Illicit TRAfficking Countermeasures Kit (EURITRACK) project, a new Tagged Neutron Inspection System (TNIS) has been developed and installed in the Port of Rijeka in Croatia. The system was based on the examination of sea containers with the 14 MeV neutron beam. During the operation the characteristic gamma rays were produced and measured by several 5''x5''x10'' NaI(Tl) detectors. During this procedure some of the detectors were exposed to an intensive neutron beam radiation. It was necessary to check for possible radiation damage of the NaI(Tl) scintillator during the gamma detector selection phase of the project. The 4''x4'' NaI(Tl) detector was exposed to 14 MeV neutrons for 20 h. From the presented results on energy resolution and activation measurements it could be concluded that there are no significant differences in energy resolution before and after the irradiation by 4.7x10(11) of 14 MeV neutrons. The only problem could be the high level of medium and long term induced activity in the energy region below 2 MeV.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.