Abstract
A 64-channel circuit Application Specific Integrated Circuit (ASIC) for charge measurement has been designed in CMOS 0.35 μm technology and characterized with electrical tests. The ASIC has been conceived to be used as a front-end for dosimetry and beam monitoring detector read-out. For that application, the circuitry is housed at a few centimeters from the irradiated area of the detectors and therefore radiation damages can affect the chip performances. The ASIC has been tested on an X-ray beam. In this paper, the results of the test and an estimate of the expected lifetime of the ASIC in a standard radio-therapeutical treatment environment are presented. An increase of the background current of 2 fA/Gy has been observed at low doses, whilst the gain changes by less than 3% when irradiated up to 15 kGy. Furthermore it has been assessed that, when used as an on-line beam monitor and the annealing effect has been taken into account, the background current increase is ∼440 fA/year.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.