Abstract
In the current study, we compared plastic matrix-assisted laser desorption/ionization (MALDI) plates pre-spotted with different solid ionic matrices. Data reflect that after 3months of storage, the standards were oxidized in α-cyano-4-hydroxycinnamic acid (HCCA) whether or not in HCCA/3-acetylpyridine (3APY) and HCCA/aniline, and certain peptides, such as ubiquitin, were not detected using the HCCA matrix, whereas they were detected in pre-spotted ionic matrices. Application in peptidomics of these MALDI matrices pre-spotted plates (after 3months of storage) with ovarian cyst fluid showed less intense signals with HCCA than with solid ionic matrices. We show that these pre-spotted ionic matrices plates can be used for relative drug quantification, high mass protein detection, and MALDI mass spectrometry imaging.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.