Abstract

A magnetic needle headspace single-drop microextraction (MN-HS-SDME) method coupled to HPLC/UV has been developed. Trihexyl(tetradecyl)phosphonium chloride was employed as an ionic liquid (IL) solvent for the headspace extraction of some chlorophenol (CP) compounds from wastewater samples. Despite of the nonmagnetic character of the IL, a significant improvement in the extraction efficiency was obtained by the magnetization of the single-drop microextraction needle using a pair of permanent disk magnets. A simplex method for the fast optimization of the experimental conditions (e.g., stirring speed, ionic strength, pH, extraction time and temperature) was used. The coefficients of determination (R2) varied between 0.9932 and 0.9989, the limits of detection were from 0.004 to 0.007μgmL-1 and the relative recoveries were in the range of 88-120% for the studied analytes. The developed MN-HS-SDME HPLC/UV method was successfully applied to the determination of CPs in industrial wastewater.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.