Abstract

An ion-trap mass spectrometer for determination of quaternary ammonium herbicides in water samples is evaluated. Implementation of the method previously optimized in a quadrupole mass spectrometer (VG Platform II) in an ion-trap (LCQ) required modifications to the sample inlet of the electrospray source. Quality parameters were established and detection limits, after trace enrichment, range 5–50 ng L−1. Good reproducibilities (day-to-day and run-to-run) <12% were obtained. The method was applied to analysis of spiked samples and the results were comparable with those determined using a quadrupole mass spectrometer.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.