Abstract

The large range of reference air kerma rates of brachytherapy sources involves the use of large-volume ionization chambers. When such ionization chambers are used the ion-recombination correction factor has to be determined. In this paper three spherical ion chambers with volumes ranging from 30 to have been irradiated by photons of a source to determine the factors. The ionization currents of the ion chambers as a function of the applied voltage and the air kerma rate have been analysed to determine the contribution of the initial and general ion recombination. The values for large-volume ionization chambers obtained by considering the general ion recombination as predominant (Almond's approach) are in disagreement with the results obtained using methods that consider both initial and general ion-recombination contributions (Niatel's approach). Such disagreement can reach 0.7% when high currents are measured for a high-activity source calibration in terms of reference air kerma rate. In this study a new `two-voltage' method, independent of the voltage ratio given by a dosimetry system, is proposed for practical dosimetry of continuous x- and gamma-radiation beams. In the case where the Almond approach is utilized, the voltage ratio should be less than 2 instead of Almond's limit of .