Abstract
An ion interaction chromatography method for the determination of nitrate and nitrite using a reversed-phase C 18 column is described. The eluent was 2.0 m M nonylammonium phosphate of pH 6.5 and the UV detection wavelength was 205 nm. For a sample volume of 200 μl detection limits of 2.0 and 1.5 μg/l were obtained for nitrite and nitrate, respectively. The change of the detection wavelength to 190 nm enables also the quantitative detection of chloride. Results obtained for natural waters have been correlated with those obtained using standard spectrophotometric procedures.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.