Abstract

Various approaches for transmitting ions from atmosphere to the deep vacuum required for mass analysis have been developed with the goal to increase the ion to gas ratio while maintaining high ion transmission efficiency. Since the vast majority of ion losses occurs in the atmospheric pressure ion source, an effective way to improve sampling of those ions is to increase the atmosphere to vacuum aperture diameter. However, as the aperture diameter is increased, the resulting intense free jet gas expansion and subsequent gas beam can scatter ions in the first vacuum region. The interface described here provides an optimized flow field to the second vacuum stage, with a unique geometry to counter the ion losses from scattering collisions with the gas. Two additional differentially pumped quadrupole ion guides are used to further improve the ion to gas ratio, resulting in an ion transfer efficiency improvement of 5-6× over a two-stage differentially pumped interface with quadrupole ion guides. The interface also demonstrates efficient declustering and fragmentation capabilities beneficial for reducing background chemical noise.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.