Abstract
There are two modes for ion energy analysis at a rf electrode in a plasma chamber. One is a dc mode in which the electric potential of the ion energy analyzer is constant in time, and another is a rf mode in which the electric potential of the analyzer oscillates with same frequency, amplitude, and phase as that of the rf electrode. For correct ion energy analysis at the rf electrode, the rf mode is necessary. For ion mass analysis in the rf mode, the electric potential of the ion mass analyzer also must oscillate. We fabricated such an ion energy and mass analyzer, and confirmed its performance.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.