Abstract

A variety of neutral carrier type ionophores for monovalent cations were employed to prepare solid-state cation-selective electrodes (SSEs) for use as a detector in single-column ion chromatography (IC). The polyurethane-based pseudoreference electrode made it possible to assemble an array type SSE detector for IC. An SSE-based detector provides not only the overall chromatogram for the separated ion species (monensin methyl ester–nonactin-based membrane), but also the enhanced chromatogram for specified ions of interest (valinomycin as K + and nonactin for NH 4 +). This feature makes it possible to perform highly quantitative analysis with low detection limits even if the separation efficiency of the ion-exchange is not sufficient. Since SSE-based IC detectors are easily miniaturized and replaceable at low cost, they are an ideal component of a portable IC system.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.