Abstract

The method is described for ion chromatographic determination of Ba, Ca, Cd, Cu, Mn, Ni, Sr, Pb and Zn. Ethylenediammonium chloride or tartrate solutions are used as eluents with a suppressor column. The detection limits for the listed elements are (in μg l−1): Ba 1.4, Ca 0.045, Cd 0.6, Cu 0.3, Mn 0.8, Ni 0.5, Pb 6.3, Sr 0.3, Zn 1.2. Relative standard deviations are 0.015–0.050 ar the 20 mg l−1 level (n = 5, p = 0.95). Interferences are considered.

Full Text
Published version (Free)