Abstract
High-performance ion chromatography (HPIC) is first successfully used to analyze tetracycline antibiotics (TCs) in this work. The TCs are well separated on a solvent compatible polymeric cation-exchange column within 12 min. Isocratic elution with acetonitrile-hydrochloride is very advantageous for routine analysis. HPIC may be seen as a specific variant of the more common high-performance liquid chromatography (HPLC) for water-soluble and polar pharmaceuticals with low hydrophobicity. The detection limits (signal-to-noise ratio=3:1) of oxytetracycline (OTC), tetracycline (TC), chlortetracycline (CTC), doxycycline (DC) are 10, 10, 20 and 20 μg l −1, respectively. Samples are prepared by vortex mixing with an ethylenediaminetetraacetic acid disodium salt (Na 2EDTA)–McIlvaine buffer (pH 4.0) solution and the mixture filtrates through a molecular weight cut-off filter. The method has been successfully applied to monitor the OTC removal rate through every reactor in the process of OTC manufacturing wastewater treatment by bio-chemical technology. It is also applicable to determine the TCs residues in milk and milk powder with satisfying results.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.