Abstract
A simple, low cost and portable iodide sensor for the indirect determination of As(III) has been developed from the polycrystalline material obtained by co-precipitation of silver sulphide and silver iodide from aqueous solution. The Ag2S-AgI sensor was found to behave well up to the iodide concentration of 1×10-6 M with a Nernstian slope 57.5 mV per decade change in concentration. The selectivity coefficient of sensor containing iodide ion in the presence of Br- and Cl- ions was found to be 1.77 × 10-2 and 7.94 × 10-3, respectively indicating that these ions do not interfere in the response of iodide sensor to iodide ion. The electrode was successfully used for the direct determination of iodide and indirect determination of As(III) by titrating with cerric ammonium sulphate at pH 2. The results obtained from this method are comparable to the existing methods for the estimation of As(III).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.