Abstract

Implants endurance as well as a good clinical tolerance depends on the recovery of a physiological stress distribution within bone after implantation. The purpose of the present work was to develop an alternative technique using Force Sensing Resistors (FSR) to gather invitro pressure values at the implant-bone interface for a cementless implant. Eight cementless femoral stems were instrumented with six calibrated FSR bonded on each facet and then implanted in eight cadaver femurs. Compression tests were performed until failure and FSR pressure values were recorded. The average failure load was 4241N. The maximum contact pressure measured with the FSR averaged 1.965MPa. FSR reached many of the requirements for an ideal implant-bone interfacial sensor. This experimentation provided invitro quantitative data on contact pressure at the implant-bone interface, which could help understanding stress shielding phenomenon and developing relevant numerical model.