Abstract
Bone drilling is a common practice of surgical treatments in orthopaedics and traumatology. Penetration of a high-speed drill into bone tissue is accompanied by generation of a significant amount of heat. Cooling of the drilling region is necessary to avoid potential risk of thermal damage to bone. The purpose of this study was to measure and predict bone temperature by conducting experiments and numerical simulations using cooling by means of irrigation at two different temperatures. A series of experiments and numerical studies were performed to investigate the effect of cooling conditions on the rise in bone temperature in drilling. The temperature increase in bone was assessed for different drilling speeds and feed rates in the presence irrigation at 5 °C and 25 °C. Bone temperature was found to be strongly affected by the drilling parameters and cooling conditions. Irrigation with water at 5 °C kept bone temperature well below the thermal threshold level. This study strongly recommends the use of irrigation at lower temperature for safe surgical incision.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Technology and health care : official journal of the European Society for Engineering and Medicine
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.