Abstract
The accuracy of serum neuron-specific enolase (NSE) measurements is critical, particularly in neurologic diseases and cancer. NSE measurements are compromised by slight, even invisible, hemolysis, which can produce apparently higher NSE levels, leading to inappropriate clinical decisions. In this article, we describe this issue and propose a solution for avoiding incorrect results. Twenty blood samples from donors with NSE values that were within the reference interval were considered. Experimental hemolysis was induced in vitro to examine the relationship between the degree of hemolysis and the increase in serum NSE. The data were then subjected to statistical analysis. There was excellent correlation (r2 0.953) between the degree of hemolysis and the rise in NSE concentration. Each hemolysis unit (equal to 1 mg/dL of free hemoglobin) corresponded to a mean value of 0.29 ± 0.09 ng/mL NSE that was released from red blood cells. The hemolysis index must be measured in every sample with no evident hemolysis before assaying it for NSE. Moreover, if the degree of hemolysis is between 5 and 30 units, the increase in NSE (from 1.5 to 9.0 ng/mL) must be calculated, and the laboratory results should be appended with comments that suggest the approximate rise in NSE.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.