Abstract

Total hip prostheses combining cobalt and titanium alloys in direct physical contact are currently being used. The purpose of this paper was to investigate the possibility of an accelerated corrosion occurring due to the coupling of these alloys. Electrochemical corrosion studies based on mixed potential and protection potential theories were used to study the in vitro effects of coupling the titanium and cobalt alloys. Verification of these studies was made by direct coupling experiments. Macroscopic examination of' surface features of four retrieved total hip prostheses ranging from 0-6 years implantation were conducted. The electrochemical studies predicted coupled corrosion potentials of -0.22 V and low coupled corrosion rates of 0.02 microA/cm2. Direct coupling experiments verified these results. Cobalt-titanium interfaces on the implants were macroscopically examined, and no instances of extensive corrosion were found. Overall, the in vitro corrosion studies and the examination of retrieved prostheses predicted no exaggerated in vivo corrosion due to the coupling of these cobalt and titanium alloys.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.