Abstract

Medical devices driven by electric power have come to be commonly used in hospitals, and rapid changes of voltage or current can easily cause them to fail. A stable and high quality power supply is indispensable in order to maintain safety in the modern clinical setting. Therefore, we investigated the quality of the power supply in a hospital and determined the tolerance of 13 pieces of medical equipment to voltage dips. The results showed little distortion of the voltage wave. However, we found an approximately 7% momentary voltage dip caused by lightening and other problems, such as 2 to 5% periodic drops in voltage and voltage wave distortions caused by incorrect grounding. In a tolerance test, the settings of some medical devices were changed at the time of automatic reboot after a disturbance. For another device, trend information was initialized.