Abstract

Palladium is a very useful chemical modifier for graphite furnace atomic absorption spectrometry. It can be used to stabilise many elements several hundred degrees higher than is possible with current methods. Its performance as a modifier is strongly affected by the sample matrix. The addition of a reducing agent provides for more consistent performance. The main purpose of the reducing agent is to “modify” the form of the palladium, guaranteeing that palladium is reduced to the metal early in the temperature programme. Differences were seen in the behaviour of the palladium modifier depending on the reduction method used. A comparison of palladium modifier methods in spike recovery studies of thallium was carried out to investigate differences between the reduction methods. It was suspected that differences in the physical form of palladium contributed to the variability in results. Scanning electron micrographs were used to investigate the physical form of palladium on the graphite surface. A method using the palladium modifier for the determination of selenium was compared with a commonly used procedure in spike recovery studies.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.