Abstract

Magnetic resonance imaging (MRI) during brachytherapy may alter the dose distribution of radioactive sources implanted in the tumor. This study investigates the impact of a magnetic field of 1.5 T, 3 T, and 7 T strengths on the dose distribution of high dose rate Co-60, Ir-192, and Yb-169, and low dose rate I-125 sources, using Geant4 Monte Carlo toolkit. After validating the simulation results by calculating the AAPM-TG43 dosimetric parameters, seven sources of each radioisotope were simulated in a water phantom, and their dose distributions were compared under the influence of a magnetic field. The simulation results indicate that using Co-60 brachytherapy under the MRI guidance is not recommended. Furthermore, the impact of a magnetic field of up to 7 T strength on the dose distribution of Ir-192, Yb-169, and I-125 sources is negligible, provided that there is no air pocket near brachytherapy sources.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.