Abstract

The objective of this work was to investigate and quantify the effect of sharp edges of the phantom on the point dose measurement during patient-specific dosimetry with Rapid Arc (RA). Ten patients with carcinoma of prostate were randomly selected for this dosimetric study. Rapid Arc plans were generated with 6 MV X-rays in the Eclipse (v 8.6.14) with single arc (clockwise). Dosimetry verification plans were generated for two phantoms (cylindrical and rectangular). The cylindrical phantom was solid water (diameter 34 cm) and the rectangular phantom was a water phantom (25 cm × 25 cm × 10 cm). These phantoms were pre-scanned in computed tomography (CT) machine with cylindrical ionization chamber (FC65) in place. The plans were delivered with Novalis Tx linear accelerator with 6 MV X-rays for both the phantoms separately. The measured dose was compared with the planned dose for both the phantoms. Mean percentage deviation between measured and planned doses was found to be 4.19 (SD 0.82) and 3.63 (SD 0.89) for cylindrical and rectangular phantoms, respectively. No significant dosimetric variation was found due to the geometry (sharp edges) of the phantom. The sharp edges of the phantom do not perturb the patient specific Rapid Arc dosimetry significantly.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.