Abstract

The aim of this study was to evaluate the errors in measurement of volumetric flow rate and wall shear rate measured in radial and ulnar arteries using a commercial ultrasound scanning system. The Womersley equations were used to estimate the flow rate and wall shear rate waveforms, based on the measured vessel diameter and centerline velocity waveform. In the experiments, each variable (vessel depth, diameter, flow rate, beam–vessel angle and different waveform) in the phantom was investigated in turn, and its value was varied within a normal range while others were fixed at their typical values. The outcomes revealed that flow rate and wall shear rate were overestimated in all cases, from around 13% to nearly 50%. It is concluded that measurements of flow rate and wall shear rate in radial and ulnar arteries with a clinical ultrasound scanner are vulnerable to overestimation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.