Abstract

The purpose of this study was to understand the PDD and OAR during electron beam therapy using lead cutout on the body surface. The Monte Carlo code PHITS version 3.24 was used to simulate PDD and OAR. The simulation results were compared with actual measurements using a silicon diode detector to evaluate the validity of the simulation results. The simulated PDD and OAR parameters of the linac agreed with the measured values within 2 mm. When the lead cutout on the body surface was used, all parameters except for R100 agreed with the measured values within 2 mm. The cutout sizes of the broad-beam square irradiation fields were 3 cm for 6 MeV, 5 cm for 12 MeV, and 8 cm for 18 MeV when the lead cutout on the body surface was used. The Monte Carlo simulation was useful for understanding the PDD and OAR of the lead cutout irradiation fields, which are difficult to measure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.