Abstract
The cooling rate to room temperature following the 400°C pre-irradiation anneal is known to affect the thermoluminescent properties of LiF:Mg,Ti (TLD-100) as a result of migration and clustering of defects during the cooling down process. In this investigation the dose response over an extended dose range from 0.01 to 7000Gy in both naturally cooled and the much slower furnace-cooled samples has been measured. Glow curve deconvolution based on first-order kinetics is employed to extract the dose response of the various glow peaks. Of especial interest is the behaviour of glow peaks 4, 5a and 5 as a function of dose. The idea is to modify the supralinear dose response of peak 5 from 1 to 30Gy to a linear behaviour. This dose range is important for clinical therapy and a linear dose response is of substantial advantage leading to both improved accuracy and precision.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.