Abstract

For low-energy beam transport of pulsed high perveance ion beams under space charge compensated conditions the knowledge of the build up time of compensation is of main interest. An LEBT system consisting of two solenoids has been set up to measure the rise time of compensation. Therefore, the beam potential has been in a residual gas ion energy analyser by time resolved measurements. The measured data has been used together with numerical simulation of different equilibrium states of the beam plasma to investigate the behavior of the compensation electrons as a function of time. The data acquired shows that the theoretical rise time of space-charge compensation is shorter by a factor of two than the build up time determined experimentally. An interpretation of the results is given.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call