Abstract

We investigated the accuracy of the recently released Remel Synergy Quad plate, a commercially available agar screening method for detecting high-level aminoglycoside and vancomycin resistance among enterococci that is based on the National Committee for Clinical Laboratory Standards recommended guidelines (National Committee for Clinical Laboratory Standards, M7-A3, 1993). The Synergy Quad correctly determined the gentamicin and streptomycin resistance status for > or = 97% of 147 Enterococcus faecalis and Enterococcus faecium isolates tested. Detection of vancomycin resistance also was reliable, as no false susceptibility occurred with 36 vancomycin-resistant E. faecalis and E. faecium strains and false resistance occurred only once with the 47 susceptible strains tested. One strain each of Enterococcus gallinarum and Enterococcus casseliflavus failed to grow on the screen, but because the true nature and significance of resistance in such isolates is unknown the implication of their screen negativity is uncertain. In summary, the Remel Synergy Quad provides a highly accurate and convenient method for susceptibility testing of enterococci against gentamicin, streptomycin, and vancomycin.