A pixel array CdZnTe imaging system, employing a 40 × 40 × 5 mm3 pixellated CdZnTe detector, is established. The imaging polarization effect in the CdZnTe pixellated detector for a collimated Cs137 Gamma source is investigated in detail. The experimental results for different irradiated fluxes indicate that excessive irradiated flux indeed causes central pixels to be shut off completely. The imaging performance of the polarized detector is severely degraded. Polarized detector counts are simultaneously reduced to one-third of the non-polarized detector counts. A theoretical model of potential distribution is also proposed by solving the Poisson equation and, in turn, the electric potential distortion for high irradiated flux is discussed by comparison with the experimental results.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.