Abstract

Today, an artificial tooth root called a dental implant is used to replace lost tooth function. Treatment with dental implants is considered an effective and safe method. However, in some cases, the use of dental implants had some failures. The success of dental implants is influenced by several biomechanical factors such as loading type, used material properties, shape and geometry of implants, quality and quantity of bone around implants, surgical method, lack of rapid and proper implant surface's integration with the jaw bone, etc. The main purpose of functional design is to investigate and control the stress distribution on dental implants to optimize their performance. Finite element analysis allows researchers to predict the stress distribution in the bone implant without the risk and cost of implant placement. In this study, the stresses created in the 3A.P.H.5 dental implant's titanium fixture and screw due to the change in abutment angles tolerance have been investigated. The results show that although the fixture and the screw's load and conditions are the same in different cases, the change of the abutment angle and the change in the stress amount also made a difference in the location of maximum stress. The 21-degree abutment puts the fixture in a more critical condition and increases the chance of early plasticization compared to other states. The results also showed that increasing the abutment angle to 24 degrees reduces the stress in the screw, but decreasing the angle to 21 degrees leads to increased screw stress and brings it closer to the fracture.