Abstract

The response of radiochromic dosimeters based on silicone is influenced by the dose-rate quenching and also the linear energy transfer (LET). However, the impact of curing time and composition on quenching has not been fully characterized. We investigated two dosimeter compositions, with 5% and 9% curing agent, which cured for 1, 3 or 5 days and that subsequently were irradiated with an 80 MeV proton beam with three different beam currents. Monte Carlo simulations were utilised for the estimation of dose, dose rate and LET profiles. We found a significant decrease in LET-dependent quenching with curing time, but no significant decrease in dose-rate-dependent quenching.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.