Abstract

One method to optimise the use of x-rays in CT and hence a reduction in patient dose is the application of automatic exposure control (AEC). This study measured the effective mAs, image noise and volume CT dose index (CTDIvol) as the result of changing the AEC index on a Siemens Somatom Definition 64 slices dual source CT scanner. The scans were performed on four phantoms of different geometries, namely the 16 and 32 cm cylindrical CTDI phantoms and two anthropomorphic phantoms, RANDO (20 cm effective diameter) and ATOM (19.8 cm effective diameter). Results showed that the effective mAs increased with increasing tube potential (kVp) and Quality Reference mAs (QRM), therefore increasing CTDIvol while reducing image noise. Meanwhile, no changes of radiation dose and image noise were observed when the pitch was increased. However, for the largest phantom (32 cm effective diameter), a constant effective mAs was found between 120 and 140 kVp. The same trend was also found with increasing QRM from 300 mAs to 400 mAs suggesting a certain limitation of the AEC has been reached. In conclusion, this study showed that AEC is affected by kVp and QRM but not by pitch selection. Further work is required to quantify the characteristics of the AEC system in relation to the mentioned parameters for better optimisation.