Abstract

Extensive studies of platinum–carbon eutectic alloy based high temperature fixed point cells have shown that this alloy has extremely good metrological potential as a temperature reference. However, it’s possible adoption as an accepted reference standard means that its eutectic temperature value will soon be agreed with an uncertainty less than most radiation thermometry scales at that temperature. Thus it will lack credibility if used as a future scale comparison artefact. To avoid this, the fixed-point cell can be deliberately doped with an impurity to change its transition temperature by an amount sufficient to test the accuracy of the scales of the institutes, involved in the comparison. In this study dopants of palladium and iridium were added to platinum–carbon to produce ternary alloy and quaternary alloy fixed-point cells. The stability of these artefacts was demonstrated and the fixed-point cells were used to compare the ITS-90 scales of NIM and NPL. It was found that the fixed point temperatures could be changed by an appreciable amount while retaining the stability and repeatability required for comparison artefacts.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.