Abstract
This paper describes the 3D dosimetry verification of a 24Gy in 3 fractions VMAT SBRT treatment delivery to the 12th thoracic vertebrae (T12). With this fractionation regime, the critical isodose value is 17.5Gy to the outer edge of the spinal cord and must not be exceeded. This is a high-risk treatment that could result in severe injury to the patient unless administered precisely. With the use of 3D normoxic polymer gel dosimetry techniques, using MRI readout, a clinical example is presented which shows that the Mid North Coast Cancer Institute’s paraspinal program is capable of delivering complex distributions with high doses to within very tight tolerances. This also demonstrates the strength of normoxic polymer gel as a 3D dosimeter.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.