Abstract

The study explored the optimum conditions for sample preparation of polyolefins for high temperature gel permeation chromatography (GPC) recommended by ASTM D 6474, and the potential for reducing solvent consumption by using columns with reduced internal diameter. A polypropylene sample with high crystallinity and high molecular weight was selected, because this is one of the most difficult polyolefins to analyze by GPC. Sequences of injections covering 20 h showed that, at 160 °C, 2 h is enough to dissolve this sample, and that the antioxidant concentration is the critical parameter in obtaining reproducible GPC results. By using one column with an internal diameter (ID) of 4.6 mm and a flow rate of 0.3 mL/min, instead of the classical set of 3 columns of 7.5 mm ID and a flow rate of 1 mL/min, the solvent consumption can be reduced to about 10%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.