Abstract

Reproducibility of quantitative PCR results is dependent on the generation of consistent calibration curves via accurate volume transfers and instrument performance. A review of 14 standard curves, using two different QuantDuo® standard DNA lots, showed variability of cycle threshold values between assays were larger than those of the Internal PCR Control (IPC). This prompted a set of experiments designed to determine the source of variability. Results showed that error introduced during DNA addition to the plate resulted in little variation. A comparison of seven independent series demonstrated cycle threshold variation between dilutions was larger than the variation expected from repeated samples. Modeling the influence of pipette errors on dilution series accuracy indicated that a more rigorous approach to external calibration curve production is required and showed that improvement in calibration curve stability is expected if the pipette conditions are carefully chosen and ⁄ or a single validated curve is utilized as the calibrator.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.