Abstract

Quality control in digital radiology can be time-consuming. Equipment self-tests may significantly decrease staff workload. The two most essential parameters for radiology systems are image quality and patient dose. Concerning patient dose, information on the dose-area product (DAP) values generally forms the basis for assessment of patient dose. DAP-values can be measured using a transmission ionisation chamber or calculated from equipment settings. In the present study, various image quality parameters were derived using a contrast-detail (C-D) phantom. The investigation included a computer-aided assessment of C-D images, which produced various parameters, and also parameters based upon scoring by human observers. In addition, another parameter was calculated from modulation transfer function (MTF) measurements. The automatically calculated parameters showed good correlation with human readings, although the number of X-ray systems studied is still limited. We propose a combined evaluation of DAP and automatically calculated C-D or MTF parameters for equipment self-tests.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.