Abstract
We report on the study of a microwave superconducting quantum interference device (SQUID) multiplexer (MW-Mux) with current noise lower than that of our typical transition-edge sensor (TES) X-ray microcalorimeter. A large mutual inductance, between SQUID loop and input inductor connected to a TES X-ray microcalorimeter in series, realizes such low current noise referred to the SQUID input, but also can cause undesirable resonances which affect the SQUID performances. In this paper, we have developed the MW-Mux with mutual inductance of about 370 pH, which is six times larger than that of the previous one, by adopting a spiral input coil longer than a half wavelength at the readout microwave frequency. Its characteristics are evaluated, and the results have shown that the distortion of the SQUID characteristics and the degradation of its noise performance are not clearly observed. Finally, we have developed an MW-Mux microstrip SQUID with a low current noise of <; 20 pA/√(Hz) which is below the typical noise of our TES X-ray microcalorimeter, 30 pA/√(Hz).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.