Abstract

The main thrust of this work is the investigation of performance of relatively new commercial LaBr3:Ce probe (Inspector 1000™ with LaBr3:Ce crystal) for gamma-ray spectroscopy and dosimetry measurements in comparison to LaCl3:Ce and NaI:Tl scintillators. The crystals were irradiated by a wide range of energies (57Co, 22Na, 18F, 137Cs and 60Co). The study involved recording of detected spectra and measurement of energy resolution, photopeak efficiency, internal radioactivity measurements as well as dose rate. The Monte Carlo package, Geant4 Application for Tomographic Emission (GATE) was used to validate the experiments. Overall results showed very good agreement between the measurements and the simulations. The LaBr3:Ce crystal has excellent energy resolution, energy resolutions of (3.37±0.05)% and (2.98±0.07)% for a 137Cs 662keV and a 60Co 1332keV gamma-ray point sources respectively, were recorded. The disadvantage of the lanthanum halide scintillators is their internal radioactivity. Inspector 1000™ with LaBr3:Ce scintillator has shown an accurate and quick dose measurements at Positron Emission Tomography (PET) Units which allows accurate assessment of the radiation dose received by staff members compared to the use of electronic personal dosimeters (EPD).

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call