Abstract

Heating and humidification control during mechanical ventilation is important. Water condensation can occur due to insufficient humidification. Condensation in the breathing circuit can cause various adverse events. In this study, we investigated whether the occurrence of tubing condensation in the breathing circuit could be affected by room temperature using a simulated respiratory circuit with a humidifier. The room temperature was set at 23°C, 25°C, and 27°C, and mechanical ventilation was performed for 8 h. The inspired gas was appropriately heated and humidified according to the manufacturer’s instructions. The weight of the circuit was measured every 2 h at room temperature to estimate the amount of condensation. During the mechanical ventilation, condensation continued to increase at a room temperature of 23°C, but only for the first 2 h at 25°C, and did not occur at 27°C. The room temperature of the indoor environment was one of the factors generating water condensation in the breathing circuit with a humidifier even when a heater wire in the breathing circuit was appropriately operated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.