Abstract
Resonance measurements and atomic force microscopy (AFM) observations were carried out by the non-contact AFM operating in various gas atmospheres (hydrogen, helium, nitrogen and argon) over the range of pressures from 0.1 to 1.1 MPa. In each atmosphere, the resonance frequency of the AFM cantilever depended on the pressure of gases studied. The plots of the relative resonance frequency at a constant pressure vs. the gas density gave a straight line. It was found that the characteristic of the resonance frequency for the AFM cantilever were dependent on the density of the gas species. The resolution of the AFM was hardly influenced by the gas atmosphere under the present experimental conditions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.